Healthcare workers need to feel confident that the personal protective equipment (PPE) they are being asked to use during an influenza pandemic will be reliable in reducing their risk of infection. Further, the equipment needs to be effective in a work environment that involves interaction with and examination of patients and long working hours in a crisis pandemic situation. As discussed in Chapter 1, PPE is one component of an overall systems approach to infection prevention and control, which during an influenza pandemic will also require environmental and policy measures including vaccination of healthcare workers, use of antiviral medications, isolation precautions, and ventilation and air exchange controls.
This chapter begins by setting out a proposed framework for the design and development of PPE for healthcare workers that will facilitate greater interaction between the end users, designers and manufacturers, and standards and certification agencies. The discussion then focuses on specific research opportunities for enhancing the current generation of PPE and concludes by identifying next steps in the design and development of PPE. The chapter’s recommendations focus on innovative and systematic approaches to the design and engineering of healthcare PPE.
A formal framework for the design and development of PPE encompasses the three phases typically associated with a product’s life cycle: user requirements analysis, design realization, and field use and evaluation.
Below are the first 10 and last 10 pages of uncorrected machine-read text (when available) of this chapter, followed by the top 30 algorithmically extracted key phrases from the chapter as a whole.
Intended to provide our own search engines and external engines with highly rich, chapter-representative searchable text on the opening pages of each chapter.
Because it is UNCORRECTED material, please consider the following text as a useful but insufficient proxy for the authoritative book pages.
Do not use for reproduction, copying, pasting, or reading; exclusively for search engines.
OCR for page 77
3
Designing and Engineering
Effective PPE
Healthcare workers need to feel confident that the personal protec-
tive equipment (PPE) they are being asked to use during an influenza
pandemic will be reliable in reducing their risk of infection. Further, the
equipment needs to be effective in a work environment that involves in-
teraction with and examination of patients and long working hours in a
crisis pandemic situation. As discussed in Chapter 1, PPE is one compo-
nent of an overall systems approach to infection prevention and control,
which during an influenza pandemic will also require environmental and
policy measures including vaccination of healthcare workers, use of anti-
viral medications, isolation precautions, and ventilation and air exchange
controls.
This chapter begins by setting out a proposed framework for the de-
sign and development of PPE for healthcare workers that will facilitate
greater interaction between the end users, designers and manufacturers,
and standards and certification agencies. The discussion then focuses on
specific research opportunities for enhancing the current generation of
PPE and concludes by identifying next steps in the design and develop-
ment of PPE. The chapter’s recommendations focus on innovative and
systematic approaches to the design and engineering of healthcare PPE.
FRAMEWORK FOR PPE DESIGN AND DEVELOPMENT
A formal framework for the design and development of PPE encom-
passes the three phases typically associated with a product’s life
cycle: user requirements analysis, design realization, and field use and
evaluation.
77
OCR for page 77
78 PREPARING FOR AN INFLUENZA PANDEMIC
Key Design Drivers
The design and development of PPE are influenced by the key factors
shown in Figure 3-1. Since meeting the regulatory standards is manda-
tory and not optional, the design and development of PPE often involve
major compromises while attempting to simultaneously achieve a maxi-
mal degree of protection with the highest level of comfort at the lowest
possible cost. For example, the degree of protection provided by protec-
tive clothing, such as a gown, can be considerably enhanced by the use
of polyethylene film without substantial additional expense, but at a sig-
nificant loss of comfort for the user. On the other hand, a high degree of
protection and comfort can be achieved, but at a much higher cost, by
using a breathable impervious nonwoven material (Pasko, 2007). Thus,
although materials and manufacturing technologies exist that can maxi-
mize any one design driver, designing the product to achieve the appro-
priate balance is ultimately dictated by the requirements of the end user.
As will be described in Chapter 4, a number of barriers and reasons
have been identified by healthcare workers regarding why they choose
not to wear PPE. These reasons include not having enough time to don
the equipment (particularly in emergency response situations), the
equipment is not available or they have not received training, the equip-
ment is uncomfortable or difficult to use, the equipment interferes with
their interaction with the patient and affects dexterity or the ability to
perform a medical procedure, or they do not see the situation as a high
risk. Better guidance is required on the unique needs of healthcare work-
ers so that appropriate performance requirements can be developed and
Degree of
Protection
Cost Comfort
Regulations
FIGURE 3-1 The design drivers for PPE.
OCR for page 77
79
DESIGNING AND ENGINEERING EFFECTIVE PPE
manufacturers can design and supply PPE to meet the specific needs of
this workforce. Moreover, since the design (or solution) space is fairly
large, it is possible to produce a large number of variations of the same
item of PPE, thereby driving up its cost. By developing PPE based on a
prescribed set of evidence-based performance requirements or standards,
manufacturers will be able to create products that will be less expensive
and more effective; such standards will also enhance compliance in
the use of PPE since they will minimize, if not eliminate, the errors
typically associated either with the selection of PPE by personnel
responsible for PPE procurement in healthcare settings or with its
use by healthcare workers themselves. Healthcare workers will be
assured that they are receiving the right level of protection in the work-
place. To realize this objective, there is a need for a structured design
and development process for PPE, as well as thorough testing and
certification efforts (Chapter 5).
User Requirements Analysis—Data Collection for Design
In the first phase of the design and development process, the re-
quirements of the end user (i.e., the healthcare worker) should be as-
sessed. The first step is to gain an understanding of the hazards and risks
associated with the use of PPE in specific environments as well as to un-
derstand the barriers to PPE use, particularly in emergency response and
crisis situations in patient care (Chapter 4). A clear understanding of the
threat will help establish the degree of protection the PPE must meet or
exceed. In the case of an influenza pandemic, this calls for an under-
standing of the nature of the influenza virus, its infectivity, and its modes
of transmission (Chapter 2). A related factor that should be considered is
the risk posed by the environment in which the healthcare worker must
operate. The continuum of risk is not clearly defined for influenza be-
cause so little is known about the routes of transmission of the virus be-
tween individuals. Further, the many unknowns concerning the nature
and level of infectivity of the influenza virus create challenges for de-
signing effective prevention measures. Unlike many industrial exposures
for which adverse health effects are the result of exposure to large
concentrations of a chemical or other hazardous agent, infectious
diseases (such as tuberculosis) may be spread by small numbers of
bacilli or viruses.
OCR for page 77
80 PREPARING FOR AN INFLUENZA PANDEMIC
As discussed in Chapter 4, research is needed that will provide a haz-
ard assessment with insights into whether specific procedures or work
situations (e.g., nebulization, endotracheal intubation, bronchoscopy,
endotracheal suctioning, cleaning patients’ rooms) place healthcare
workers at higher levels of risk of influenza infection. According to the
2006 interim guidelines for an influenza pandemic, N95 respirators are
recommended for healthcare workers in caring for patients with con-
firmed or suspected influenza or in situations, such as bronchoscopy or
resuscitation, that are likely to generate infectious respiratory aerosols
(CDC, 2006). McCullough and Brosseau (1999) present a qualitative
framework for the selection of respirators for the control of worker expo-
sure to infectious aerosols, especially in situations where information on
occupational exposure limits, toxicity, and airborne concentrations is
absent. As stated in Chapter 2, information on the modes of transmission
of the influenza virus is scarce and this type of qualitative approach may
be valuable in assessing the risk in the healthcare setting during a pan-
demic. The authors urge that assessments be conducted by industrial hy-
gienists or other trained professionals.
In developing evidence-based performance requirements, the ideal
data acquisition process would involve use of the PPE component in the
field and assessing the requirements; however, in the event this is not
feasible, the data acquisition process should, at the very least, simulate
the real-world usage of the specific component of the PPE ensemble. For
instance, the healthcare worker will sweat during the course of normal
day-to-day activities, and this in turn will affect the performance of the
PPE—the respirator may change its position on the user’s face or the
gown may become increasingly uncomfortable if it does not effectively
wick away perspiration from the user’s skin. Therefore, a treadmill or
similar method can be used to simulate the use of PPE components to
better understand and determine their performance requirements.
The next step is to identify the key characteristics that should be con-
sidered in the design of the PPE component. As shown in Figure 3-2,
these involve considerations of function, use, comfort and wearability,
durability, maintenance and reuse, aesthetics, and cost.
OCR for page 77
81
DESIGNING AND ENGINEERING EFFECTIVE PPE
Evidence-Based Performance Requirements
Functionality Usability Comfort and Wearability Durability
• Protect against • Maintain biomechanical • Comfortable—no skin • Adequate wear life
• Strength—tear,
influenza virus efficiency and sense of touch irritation or pressure
• Guard against and feel points tensile, burst
• Odor-free • Abrasion resistance
• Prolonged use
contact with
• Hypoallergenic • Corrosion
contaminated without discomfort
• Accommodate wide range of • Breathable—air resistance
fluids and
aerosols users (face and body profiles) permeable
• Compatability across various • Moisture absorbent—
elements of the PPE wickability
• Low bulk and weight
ensemble and with other
• Dimensional stabiltiy
equipment (e.g., stethoscope)
• Non-startling to patients and • Easy to put on and
families take off (don and doff)
• Facilitates communication with
others (verbal, facial)
Cost
• Product cost
Maintenance and Aesthetics
• Total life-cycle
Reuse
• Variety of styles cost
• Easy to • Minimal environ-
and colors
• Customizable
decontaminate and mental impact
discard disposable
elements
• Easy to clean and
replace parts in
reusable PPE
FIGURE 3-2 A structured approach to evidence-based performance
requirements.
For example, protection against the influenza virus and guarding
against splashes and contact with bodily fluids are the major functional
requirements of PPE. However functionally effective the PPE may be, it
is unlikely to be used regularly in the field if the efficiency of the user in
carrying out his or her task is impaired by the PPE. PPE should not affect
the biomechanical efficiency (work and energy) of healthcare workers,
especially since they rely on extensive interaction with the patient and
must be able to hear the patient’s respiration and heartbeat, touch and
feel the patient’s body, and so on. The PPE should be odor-free and hy-
poallergenic and should comfortably fit a variety of body forms includ-
ing facial profiles. Its appearance should not startle patients, especially
younger children. It should also facilitate verbal and facial communica-
tion with patients. User instructions that accompany PPE products should
clearly specify appropriate practices to promote their correct usage. In
terms of comfort and wearability, the PPE should be comfortable to wear
during work activities and should not have any pressure points or cause
skin irritation. It should be breathable and have good moisture absorp-
OCR for page 77
82 PREPARING FOR AN INFLUENZA PANDEMIC
tion. It should be lightweight and have excellent dimensional stability
since it will be subjected to extensive stresses and strains during wear. It
should be easy to put on and take off (don and doff), especially in a very
short period of time. It should be durable, with the wear life depending
on the type of ensemble (e.g., gown, respirator), and should be of sound
construction to prevent or minimize damage due to tear, tensile, and
puncture deformations. Careful consideration should be given to the
trade-offs between disposable and reusable PPE, particularly given the
extreme demands that would be placed on a disposable PPE supply in an
influenza pandemic. Maintenance and reuse are key factors for consid-
eration in developing performance requirements (IOM, 2006). Minimiz-
ing the environmental impact of PPE cleaning or discard should also be
considered. The PPE should be customizable to meet the wearer’s aes-
thetic needs including those of style and color. Finally, the product cost
and the total life-cycle cost should be specified as part of the require-
ments analysis. A similar user requirements analysis process has been
employed successfully in the design and development of the Wearable
Motherboard or Smart Shirt, an intelligent garment for biomedical moni-
toring (Rajamanickam et al., 1998; Park and Jayaraman, 2003).
Design Realization—Design and Engineering
The second step in the framework is realization of the design by
translating the evidence-based performance requirements into the spe-
cific design of the PPE component in light of the regulatory requirements
as shown in Figure 3-3.
This part of the process begins with making appropriate trade-offs
between the design drivers of degree of protection, comfort, and cost
for the specific PPE component being designed. Once this “degree
of protection-comfort-cost” solution space has been established, appro-
priate materials and manufacturing processes need to be chosen. For
example, the level of required filter efficiency will determine the choice
of materials and specific treatments during the manufacturing process for
a respirator. Similarly, appropriate finishing treatments should be chosen
OCR for page 77
83
DESIGNING AND ENGINEERING EFFECTIVE PPE
Evidence-based Manufacturing processes
performance requirements
FMEA
Degree of protection- Field-testing of
comfort- cost solution space PPE component
Regulations Materials
Determines the
Define the Results in the
selection of
Failure modes
and effects
analysis
FIGURE 3-3 PPE design life cycle: evidence-based performance require-
ments through field testing.
to provide the required degree of thermal comfort for gowns to ensure
the comfort of the healthcare workers who are using them. The potential
modes of failure of the PPE component in the field should be anticipated
and the product suitably designed to guard against such failures. A for-
mal failure modes and effects analysis process should be adopted to en-
sure the robustness of the resulting design. This process is aimed at
proactively identifying where and how equipment and processes might
fail and focusing on where changes are needed (IHI, 2007).
Field Use and Evaluation: Product in Use
As shown in Figure 3-3, and discussed in Chapter 5, in the final
phase of the framework, the developed PPE component should be tested
and evaluated in the field for a realistic assessment of its performance
and to monitor any unintended consequences of use. For respirators, this
will necessitate the integration of field testing into the certification proc-
ess. During this field testing, the product should be subjected to the vari-
ous failure modes identified earlier as part of the FMEA process.
Protocols should be put in place to obtain feedback from users during the
testing, and these inputs should be used to refine and enhance the design.
For example, an ongoing study of the tolerability of various respirators
and respirator-mask configurations should provide valuable insights into
real-world usage (Radonovich, 2007).
Once the product has been placed in service, appropriate mechanisms
should be established to obtain continuous feedback on its performance.
Programs should be instituted to ensure compliance with the right use of
the right PPE for the right level of risk.
OCR for page 77
84 PREPARING FOR AN INFLUENZA PANDEMIC
In summary, the proposed formal PPE development framework calls
for a greater degree of input and collaboration between the various
stakeholders associated with PPE—the users (i.e., healthcare workers),
the designers and manufacturers, and the regulatory or certification
agencies (i.e., the National Institute for Occupational Safety and Health
[NIOSH], the Food and Drug Administration [FDA], and the Occupa-
tional Safety and Health Administration [OSHA]) responsible for certify-
ing and approving PPE. Such a systems and iterative approach will lead
to the development and deployment of effective and wearable PPE that
can be used in the range of healthcare settings from patients’ homes to
hospitals to long-term care facilities. The remainder of this chapter
identifies a set of research opportunities to enhance the current genera-
tion of PPE and spur the innovations that will result in a new generation
of protective equipment.
RESPIRATORY PROTECTION: RESEARCH NEEDS
The fundamental principle for making decisions regarding the selec-
tion and use of respiratory protection is to understand the nature of the
hazard and the risks that the wearer is expected to encounter when wear-
ing that protection. While there is extensive knowledge regarding the
efficacy of respiratory protection, little is known about the extent to
which aerosol transmission contributes to the overall risk of infection by
the influenza virus. Therefore, the most critical research need regarding
respiratory protection for healthcare workers, as discussed in Chapter 2,
is accurately defining the modes of transmission of the influenza virus
and the likelihood of infection by each route. Lacking this knowledge,
the selection and use of appropriate respiratory protection is qualitative
and subject to opinions regarding acceptability of risk.
Respiratory protection will be necessary in an influenza pandemic if
there is a likelihood of aerosol transmission. If properly selected and
used, respiratory protection has been demonstrated to significantly re-
duce hazardous exposures. However, much of this work has been con-
ducted in industrial settings and has focused on chemical exposures.
When compared to no respiratory protection, Barnhart and colleagues
(1997) estimated that the use of respiratory protection reduces risks of
skin test conversion for tuberculosis by the following proportions: surgi-
cal mask, 2.4-fold; disposable dust, fume, mist, or high-efficiency par-
ticulate air filtering (HEPA) mask, 17.5-fold; elastomeric HEPA
OCR for page 77
85
DESIGNING AND ENGINEERING EFFECTIVE PPE
cartridge respirator, 45.5-fold; or powered air-purifying respirator
(PAPR),1 238-fold. Teleman and colleagues (2004) found that the consis-
tent use of N95 filtering facepiece respirators by healthcare workers for
contact with severe acute respiratory syndrome (SARS) patients was
strongly protective regarding risk of SARS infection (OR [odds ratio]
0.1, 95% CI [confidence interval] 0.02 to 0.86). A limited number of
studies have looked at the effectiveness of PPE in other infectious dis-
ease situations (Table 1-4).
As discussed in Chapter 1, NIOSH has authority to define the con-
struction and performance of respirators and to certify respirators for use
that meet those requirements (NIOSH, 2004a); OSHA regulates the use
of respirators in the workplace (OSHA, 1998). The FDA has
regulatory authority to provide manufacturers with the approval
to market respirators and other PPE (e.g., gowns, gloves) that will
be used in patient care. Additionally, the American National Standards
Institute (ANSI) has issued a consensus standard on the use of respira-
tory equipment that is relevant to the use of respirators in the healthcare
setting (ANSI, 2001). Respirators approved by NIOSH for protection
from aerosols are broadly categorized by whether they are air purifying
or air supplying.2
The types of respirators that have been designated for use against in-
fluenza (CDC, 2006; OSHA, 2007b) are negative-pressure3 air-purifying
respirators or PAPRs. For negative-pressure air-purifying respirators, the
level of protection from aerosol exposure is primarily a function of leak-
age through the faceseal due to the negative pressure created inside the
facepiece of these respirators when the wearer inhales. Penetration may
also occur through the respirator filter media. For a PAPR, the level of
protection is primarily a function of the flow rate of air into the facepiece
and secondarily of the efficiency of the filter. Respirators worn by
healthcare workers will not only protect them, but may also reduce the
spread of disease from one patient to another (via the healthcare worker)
or from an infected but asymptomatic healthcare worker. Determining
whether exhaled air from workers needs to be filtered is a critical re-
1
In this report, the term PAPR is used to refer to loose-fitting devices unless otherwise
specified.
2
Air-purifying respirators use a filter, cartridge, or canister to remove air contaminants
(ambient air passes through the air-purifying element). Air-supplying respirators supply
the user with breathable air from a source independent of the ambient air (OSHA, 2007a).
3
Air pressure inside the facepiece during inhalation is lower than the ambient air pres-
sure; this allows air to flow through the filter and into the facepiece.
OCR for page 77
86 PREPARING FOR AN INFLUENZA PANDEMIC
search item. For filtering facepieces, this is accomplished by the elimina-
tion of an exhalation valve, but there is no current solution for PAPRs.
Medical masks are not designed to offer respiratory protection to the
wearer (Chapter 1). These masks protect patients from droplets in the
wearer’s exhaled breath and are not intended to fit tightly on the wearer’s
face or to be constructed of high-efficiency filter media. Medical masks
may serve to provide a barrier to infectious droplets but are not consid-
ered respiratory protection. In the aftermath of the SARS outbreaks, re-
searchers have conducted several studies to examine the level of
protection that medical masks may provide to the wearer (for example,
Balazy et al., 2006b; Li et al., 2006b). Further research is needed to clar-
ify the role of medical masks in providing barrier protection during an
influenza pandemic as these masks are widely available and will be ac-
cessible to healthcare workers and to the general public.
Enhancing the Fit
Faceseal leakage is the most critical factor in the ability of a respira-
tor to protect the wearer from exposure to airborne contaminants. OSHA
requires that respirators be qualitatively or quantitatively fit tested before
they are used (OSHA, 1998). Three important components of a respira-
tory protection program are selecting the right size and shape for the
wearer’s face, confirming fit by testing, and proper and consistent use of
the respirator when worn. The fit factor (FF) is the fundamental parame-
ter describing the effectiveness of the quality of the seal between the res-
pirator and the wearer’s face and is defined as the reciprocal of the
fraction of the contaminant concentration entering a respirator through
leaks. The fit factor is measured and determined by fit testing, which can
be conducted using quantitative or qualitative methods. Qualitative
methods rely on the wearer to detect the presence of the challenge agent
inside the respirator by smell or taste. Quantitative testing methods
measure the amount of leakage of the contaminant into the respirator
facepiece and include test aerosol, ambient aerosol, and dynamic nega-
tive-pressure tests. Filtering facepiece and half-mask respirators can be
tested by both methods. However, full-face respirators and tight-fitting
PAPRs must be tested by quantitative methods.
Fit testing and training on how to don, wear, and doff a respirator
have been shown to increase the protection provided by the respirator
while in use. One-on-one and classroom training significantly increase
fit test pass rates compared to no training at all (Hannum et al., 1996).
OCR for page 77
87
DESIGNING AND ENGINEERING EFFECTIVE PPE
After conducting aerosol ventilation studies using technetium-99 (99mTc),
Huff and colleagues (1994) found that personnel wearing fit-tested respi-
rators had significantly lower counts from radiation contamination (disin-
tegrations per minute) on nasal swabs than those wearing respirators
that were not fit tested or medical masks. Other studies have shown that
fit testing increases simulated workplace protection factors for
elastomeric and filtering facepiece respirators (Coffey et al., 1999, 2004;
Lawrence et al., 2006). In focus group discussions, healthcare workers
during the SARS outbreaks expressed concerns about the variability be-
tween fit testing and training methods used by different healthcare facili-
ties (Yassi et al., 2004). Increased standardization of fit testing and
training methods should be explored as should simpler, more efficient
methods of fit testing.
In use, the efficacy of the faceseal can vary greatly and may not nec-
essarily be related to the fit factor as determined by fit testing. The mini-
mum acceptable level of fit under these use conditions is the assigned
protection factor (APF). The APF is defined as the anticipated level of
protection provided by the respirator (based on supplying properly fitted
and functioning respirators to a given percentage of trained users)
(Bollinger, 2004). APFs are based on the analysis of workplace protec-
tion factor (WPF) and simulated workplace protection factor (SWPF)
studies (Coffey et al., 2004); the higher the APF value, the greater is the
expected level of respiratory protection. OSHA, NIOSH, and ANSI have
defined APFs for classes of respirators based on facepiece type and res-
piratory inlet covering (Table 3-1; OSHA, 2006).
The actual level of protection provided by respirators when worn un-
der various work conditions is measured by the total inward leakage
(TIL). This is the sum of the leakage through filters, respirator compo-
nents (exhalation valves), and faceseals—faceseal leakage being the
most critical and variable factor. The TIL for various models within a
respirator class or type has been shown to vary significantly, and some
models have measured penetration values greater than 10 percent. Coffey
and colleagues (1999) found significant variation in SWPFs of 21 N95
filtering facepiece respirators. Similarly, a study of 18 N95 filtering
facepiece respirators found that 5th percentile SWPFs without fit testing
ranged from 1.3 (indicating virtually no protection) to 48.0; fit testing
was found to increase protection (Coffey et al., 2004). A TIL study using
a standard European test method (EN 13274-1) found that half-mask
elastomeric facepiece respirators had less leakage than filtering facepiece
respirators and that leakage was significantly different between classes of
OCR for page 77
102 PREPARING FOR AN INFLUENZA PANDEMIC
Currently, fit testing is a critical requirement for ensuring the efficacy
of respirators. However, it may not always be complied with for various
reasons, including time and cost. Moreover, current methods preclude the
fit testing of individuals with facial hair, and respirators are not designed
specifically for young children. Therefore, the use of shape memory
polymers in the design of respirators should be investigated to enhance
ease of fit and comfort and potentially to minimize fit testing. A short-
term goal should be to develop respirators that would be easy to fit, while
a longer-term goal should be to find a way to obviate the need for fit test-
ing of respirators while being efficacious for all individuals.
Chemical Treatments on PPE with Biocidal Properties
There are times when the fit of a respirator is compromised and
pathogens can gain entry to the face (Li et al., 2006a). Moreover, the pro-
tective effect of N95 respirators and medical masks is maintained only
when the surface layer is hydrophobic and dry. Therefore, when the PPE
is wetted, protection is reduced significantly. Also, if the surface is con-
taminated with infectious agents, pathogens may penetrate the protective
layers along with the droplets.
The use of biocidal compounds as coatings for PPE is being explored
(Sun and Xu, 1998; Li et al., 2006a). Li and colleagues (2006a) have de-
veloped an antimicrobial nanoparticle coating from a mixture of silver
nitrate and titanium dioxide and demonstrated its effectiveness against
common hospital pathogens. In addition, Baker and colleagues (2005)
have demonstrated that complete cytotoxicity to bacteria cells was possi-
ble at low concentrations of silver nanoparticles. These promising studies
highlight the value of such finishing treatments in enhancing the protec-
tion afforded by PPE to healthcare workers. Therefore, research should
be directed to investigate the use of chemical treatments (e.g., using
nanoparticles) to impart biocidal properties to PPE to enhance their pro-
tection capability and possibly extend their useful life. User safety is the
primary consideration; testing standards will be needed to ensure that
biocidal materials do not pose hazards to PPE wearers.
OCR for page 77
103
DESIGNING AND ENGINEERING EFFECTIVE PPE
OPPORTUNITIES FOR ACTION
As discussed throughout this chapter, there are a number of areas in
which research is needed to improve the wearability, functionality, and
other critical aspects of healthcare PPE. The committee has identified
several key actions that if addressed expeditiously (in the next 6 to 12
months) could have a significant impact on improving the nation’s readi-
ness for pandemic influenza; longer-term opportunities and research
questions abound and need to be explored for improving healthcare PPE
products so that they can be used more effectively, with greater ease and
comfort, and for longer periods of time.
Immediate Opportunities
There is an immediate need to examine the design of PPE for health-
care workers, to improve coordination and expedite approval, and to un-
derstand the efficacy of various decontamination techniques (e.g., bleach,
microwave radiation, ultraviolet light) that could be employed on PPE in
a healthcare setting. Questions of interest include the following:
• For what period of time does PPE remain contaminated with in-
fectious influenza viruses, and what improvements can be made in doff-
ing and decontamination procedures given that information? What are
the appropriate PPE decontamination strategies that would not compro-
mise the integrity of the PPE while being easy and cost-effective to im-
plement in a healthcare setting?
• What are the differences in protection of N95 versus N100 or
other respirators if exposed to human and avian influenza aerosols?
• Current PAPRs are designed to provide extremely high flow
rates to protect the worker in an industrial setting. While appropriate to
protect from significant dust exposures, they present serious design im-
pediments for the healthcare worker. What are the flow rates and maxi-
mum noise levels that would be required for NIOSH to certify a PAPR
that would provide adequate protection for healthcare workers? What is
the risk to patients from healthcare workers wearing PAPRs (from unfil-
tered exhaled air), and what design modifications would be needed to
eliminate such risk as well as facilitate interactions with patients?
• Could a nondisposable respirator be designed that could be eas-
ily decontaminated and cost-effective?
OCR for page 77
104 PREPARING FOR AN INFLUENZA PANDEMIC
• What immediate systemic or strategic measures can be taken to
facilitate closer collaboration between healthcare workers (end users),
PPE manufacturers, and certification or regulatory agencies on the design
and development of PPE for healthcare?
Long-Term Key Research Needs
• What protective roles do gloves, gowns, and face shields or other
eye protection play in preventing influenza transmission? What protec-
tion would medical masks provide to the wearer during an influenza
pandemic?
• Do specific procedures (e.g., nebulization, endotracheal intuba-
tion, bronchoscopy, cleaning of patients’ rooms) place healthcare work-
ers at higher levels of risk of influenza infection? To what extent
do various types of PPE offer protection during these procedures and
processes?
• What technologies can improve fit to circumvent the need for fit
testing?
• What innovative designs can improve wearability issues regard-
ing PPE?
• Can the protection levels of the PPE worn by healthcare workers
(e.g., N95 respirators) be continuously monitored during use to provide
an alert to change the PPE when it is no longer effective?
• How does the penetration risk of N95 respirators made of differ-
ent materials and designs change with high inhalation rates?
• How does the level of protection afforded by N95 change with
and without fit testing?
• What is the impact of masking influenza patients on transmission
risk? If effective, how long before the medical mask needs to be
changed?
• What are the best practices for PPE removal to minimize risk of
self-inoculation?
• What are the risks of self-inoculation when changing PPE (i.e., is
the true acquisition risk the same when wearing a medical mask and
changing to an N95 for high-risk procedures versus wearing an N95
throughout the shift)?
OCR for page 77
105
DESIGNING AND ENGINEERING EFFECTIVE PPE
SUMMARY AND RECOMMENDATIONS
Healthcare workers need PPE that provides protection against influ-
enza transmission and that can be worn while working without adding
undue physiological burdens. Designing and engineering effective PPE
that will meet the needs of healthcare workers during an influenza pan-
demic will require careful consideration of three key factors: protection,
cost, and comfort, while also achieving certification and approval criteria
established by FDA, NIOSH, and other relevant agencies and organiza-
tions. Critical to the design and development of PPE are a more thorough
understanding of the threats posed by the influenza virus (see Chapter 2)
and greater engagement of healthcare workers in the design and testing
processes to provide information on the risks and the workplace envi-
ronment. Innovative designs and materials are needed for the next gen-
eration of PPE for healthcare workers. For respirators, the filter and the
faceseal are the critical issues; other types of PPE provide barrier protec-
tion and require innovations particularly regarding the interface between
PPE (e.g., between eye protection and respirators).
The development of design and performance standards is envisioned
as an iterative process that will lead to more effective and wearable PPE
products based on evolving technologies and feedback from all stake-
holders including data from researchers on the transmission of influenza
and input from healthcare workers on performance requirements. Based
on an in-depth analysis of the design and engineering of effective PPE
for healthcare workers, the committee has developed the following set of
recommendations:
Recommendation 2 Define Evidence-Based Performance Re-
quirements (Prescriptive Standards) for PPE
NIOSH, through the National Personal Protective Technology
Laboratory (NPPTL), in collaboration with extramural re-
searchers, manufacturers, and regulatory agencies, should de-
fine a set of evidence-based performance requirements or
prescriptive standards for PPE to facilitate their design and
development that optimally balances the cost, comfort, and
degree of protection of PPE and enhances compliance with
their use in the field.
OCR for page 77
106 PREPARING FOR AN INFLUENZA PANDEMIC
Recommendation 3 Adopt a Systems Approach to the Design
and Development of PPE
NIOSH should promote a systems approach to the design, de-
velopment, testing, and certification of PPE using evidence-
based performance requirements or prescriptive standards
and fostering closer collaboration between users, manufac-
turers, and research and regulatory agencies.
Recommendation 4 Increase Research on the Design and En-
gineering of the Next Generation of PPE
NIOSH, the Department of Homeland Security, the Depart-
ment of Defense, manufacturers, and other relevant organiza-
tions and agencies should fund research directed at the design
and development of the next generation of respirators, gowns,
gloves, and eye protection for healthcare workers that would
enhance their safety and comfort by
• utilizing innovations in materials such as shape mem-
ory polymers (e.g., to obviate fit testing and enhance
fit of respirators and comfort of gowns) and finishing
treatments (e.g., safe antimicrobial or biocidal
finishes);
• developing more effective and consistent faceseals for
respirators, including examination of the effect of
wear and repeated donning and doffing on the qual-
ity of the faceseal of filtering facepiece respirators,
and research on the effect of respirator filter effi-
ciency on faceseal leakage and degree of protection;
• providing a seamless interface between PPE (e.g., eye
protection and respirators);
• designing respirator facepieces to integrate medical
devices such as a stethoscope and to improve commu-
nication between the user and others;
• establishing a new set of performance requirements
for PAPRs and for reusable filtering facepiece respi-
rators that meet the needs of healthcare workers; and
• incorporating sensors into PPE to detect breaches
and notify users of end of service life and other pro-
tection information.
OCR for page 77
107
DESIGNING AND ENGINEERING EFFECTIVE PPE
Recommendation 5 Establish Measures to Assess and Compare
the Effectiveness of PPE
NIOSH, through NPPTL, should develop and promote a vali-
dated set of measures for comparing the effectiveness of PPE
products. The goal is a set of measures that would allow
users to compare and select appropriate PPE commensurate
with the assessed risk and desired level of protection. Particu-
lar attention should be paid to disseminating information
to healthcare workers on PPE effectiveness relevant to
influenza.
These efforts require:
• expedited efforts to finalize a standardized method for
measuring the total inward leakage of respirators as
part of the NIOSH respirator approval protocols;
• clear measures of filter efficiency; and
• clear measures for comparing the effectiveness of
respirators, gowns, gloves, eye protection, and other
types of PPE based on evidence-based performance
requirements.
REFERENCES
ANSI (American National Standards Institute). 2001. ANSI Z88.2—American
national standard for respiratory protection. Washington, DC: ANSI.
Baker, C., A. Pradhan, L. Pakstis, D. J. Pochan, and S. I. Shah. 2005. Synthesis
and antibacterial properties of silver nanoparticles. Journal of Nanoscience
and Nanotechnology 5(2):244-249.
Balazy, A., M. Toivola, T. Reponen, A. Podgorski, A. Zimmer, and S. A.
Grinshpun. 2006a. Manikin-based performance evaluation of N95 filtering-
facepiece respirators challenged with nanoparticles. Annals of Occupational
Hygiene 50(3):259-269.
Balazy, A., M. Toivola, A. Adhikari, S. K. Sivasubramani, T. Reponen, and S.
A. Grinshpun. 2006b. Do N95 respirators provide 95% protection level
against airborne viruses, and how adequate are surgical masks? American
Journal of Infection Control 34(2):51-57.
Barnhart, S., L. Sheppard, N. Beaudet, B. Stover, and J. Balmes. 1997. Tubercu-
losis in health care settings and the estimated benefits of engineering controls
and respiratory protection. Journal of Occupational and Environmental
Medicine 39(9):849-854.
OCR for page 77
108 PREPARING FOR AN INFLUENZA PANDEMIC
Bentley, M. 1996. Blood and body fluid exposures to healthcare workers’ eyes
while wearing faceshields or goggles. Advances in Exposure Prevention
2(4):9-11.
Bollinger, N. 2004. NIOSH respirator selection logic. National Institute for Oc-
cupational Safety and Health, DHHS-NIOSH Publication No. 2005-100.
http://www.cdc.gov/niosh/docs/2005-100/pdfs/05-100.pdf (accessed June 20,
2007).
Brennan, M. 2001. Suite of shape memory polymers. Chemical & Engineering
News 79(6):5. http://pubs.acs.org/cen/topstory/7906/7906notw1.html (ac-
cessed June 18, 2007).
Brosseau, L. M., N. V. McCullough, and D. Vesley. 1997. Mycobacterial aero-
sol collection efficiency of respirator and surgical mask filters under varying
conditions of flow and humidity. Applied Occupational & Environmental
Hygiene 12(6):435-445.
Campbell, D. 1984.The theoretical effect of filter resistance and filter penetra-
tion on respirator protection factors. Journal of the International Society for
Respiratory Protection 2:198-204.
CDC (Centers for Disease Control and Prevention). 1987. Update: Human im-
munodeficiency virus infections in health-care workers exposed to blood of
infected patients. Morbidity and Mortality Weekly Report 36(19):285-289.
CDC. 2006. Interim guidance on planning for the use of surgical masks and
respirators in health care settings during an influenza pandemic.
http://www.pandemicflu.gov/plan/healthcare/maskguidancehc.html (accessed
June 19, 2007).
Chen, C. C., M. Lehtimaki, and K. Willeke. 1993. Loading and filtration charac-
teristics of filtering facepieces. American Industrial Hygiene Association
Journal 54(2):51-60.
Coffey, C. C., D. L. Campbell, and Z. Zhuang. 1999. Simulated workplace per-
formance of N95 respirators. American Industrial Hygiene Association Jour-
nal 60(5):618-624.
Coffey, C. C., R. B. Lawrence, D. L. Campbell, Z. Zhuang, C. A. Calvert, and P.
A. Jensen. 2004. Fitting characteristics of eighteen N95 filtering-facepiece
respirators. Journal of Occupational and Environmental Hygiene 1(4):262-
271. Erratum, Journal of Occupational and Environmental Hygiene 1(8):558.
DHHS (U.S. Department of Health and Human Services). 2005. HHS pandemic
influenza plan. http://www.hhs.gov/pandemicflu/plan/ (accessed June 20,
2007).
European Standards. 2001. EN 13274-1. Respiratory protective devices—
Methods of test-part I: Determination of inward leakage and total inward
leakage, British Standard.
FDA. 2006. Medical devices; patient examination and surgeons’ gloves; test
procedures and acceptance criteria. Final rule. Federal Register
71(243):75865-75879.
OCR for page 77
109
DESIGNING AND ENGINEERING EFFECTIVE PPE
Frund, Z. 2007. Next step in materials engineering. Presentation at the Institute
of Medicine Workshop on Personal Protective Equipment for Healthcare
Workers in the Event of Pandemic Influenza: Next Steps and Research Direc-
tions, February 22, 2007. Washington, DC.
Garner, J. S., and HICPAC (Hospital Infection Control Practices Advisory
Committee). 1996. Guideline for isolation precautions in hospitals. Infection
Control and Hospital Epidemiology 17(1):53-80.
Granzow, J. W., J. W. Smith, R. L. Nichols, R. S. Waterman, and A. C. Muzik.
1998. Evaluation of the protective value of hospital gowns against blood
strike-through and methicillin-resistant Staphylococcus aureus penetration.
American Journal of Infection Control 26(2):85-93.
Han, D. H., and J. Lee. 2005. Evaluation of particulate filtering respirators using
inward leakage (IL) or total inward leakage (TIL) testing—Korean experi-
ence. Annals of Occupational Hygiene 49(7):569-574.
Hannum, D., K. Cycan, L. Jones, M. Stewart, S. Morris, S. M. Markowitz, and
E. S. Wong. 1996. The effect of respirator training on the ability of healthcare
workers to pass a qualitative fit test. Infection Control and Hospital Epidemi-
ology 17(10):636-640.
Hinds, W. 1999. Aerosol technology: Properties, behaviors, and measurement
of airborne particles, 2nd ed. New York: John Wiley & Sons.
Hosoglu, S., M. K. Celen, S. Akalin, M. F. Geyik, Y. Soyoral, and I. H. Kara.
2003. Transmission of hepatitis C by blood splash into conjunctiva in a nurse.
American Journal of Infection Control 31(8):502-504.
Huff, R. D., P. Horwitz, and S. J. Klash. 1994. Personnel protection during aero-
sol ventilation studies using radioactive technetium (Tc99m). American In-
dustrial Hygiene Association Journal 55(12):1144-1148.
IHI (Institute for Healthcare Improvement). 2007. Failure modes and effects
analysis tool. http:www.ihi.org/ihi/workspace/tools/fmea (accessed June 20,
2007).
IOM (Institute of Medicine). 2006. Reusability of facemasks during an influenza
pandemic. Washington, DC: The National Academies Press.
Ippolito, G., V. Puro, N. Petrosillo, G. De Carli, G. Micheloni, and E. Magliano.
1998. Simultaneous infection with HIV and hepatitis C virus following occu-
pational conjunctival blood exposure. Journal of the American Medical Asso-
ciation 280(1):28.
Janssen, L., and R. Weber. 2005. The effect of pressure drop on respirator
faceseal leakage. Journal of Occupational and Environmental Hygiene
2(7):335-340.
Kanaoka, C., H. Emi, and T. Ishiguro. 1984. Time dependency of collection
performance of electret filters. In Aerosols, edited by B. Y. H. Liu, D. Y. H.
Pui, and H. J. Fissan. New York: Elsevier Science Publishing. Pp. 613-616.
OCR for page 77
110 PREPARING FOR AN INFLUENZA PANDEMIC
Krishnan, U., K. Willeke, A. Juozaitis, T. Myojo, G. Talaska, and R. Shukla.
1994. Variation in quantitative respirator fit factors due to fluctuations in leak
size during fit testing. American Industrial Hygiene Association Journal
55(4):309-314.
Lawrence, R. B., M. G. Duling, C. A. Calvert, and C. C. Coffey. 2006. Com-
parison of performance of three different types of respiratory protection de-
vices. Journal of Occupational and Environmental Hygiene 3(9):465-474.
Lee, K., A. Slavcev, and M. Nicas. 2004. Respiratory protection against Myco-
bacterium tuberculosis: Quantitative fit test outcomes for five type N95 filter-
ing-facepiece respirators. Journal of Occupational and Environmental
Hygiene 1(1):22-28.
Leonas, K. K., and R. S. Jinkins. 1997. The relationship of selected fabric char-
acteristics and the barrier effectiveness of surgical gown fabrics. American
Journal of Infection Control 25(1):16-23.
Li, Y., P. Leung, L. Yao, Q. W. Song, and E. Newton. 2006a. Antimicrobial
effect of surgical masks coated with nanoparticles. Journal of Hospital Infec-
tion 62(1):58-63.
Li, Y., T. Wong, J. Chung, Y. P. Guo, J. Y. Hu, Y. T. Guan, L. Yao, Q. W.
Song, and E. Newton. 2006b. In vivo protective performance of N95 respira-
tor and surgical facemask. American Journal of Industrial Medicine 49:1056-
1065.
McCullough, N. V., and L. M. Brosseau. 1999. Selecting respirators for control
of worker exposure to infectious aerosols. Infection Control and Hospital
Epidemiology 20(2):136-144.
McCullough, N. V., L. M. Brosseau, and D. Vesley. 1997. Collection of three
bacterial aerosols by respirator and surgical mask filters under varying condi-
tions of flow and relative humidity. Annals of Occupational Hygiene
41(6):677-690.
Moyer, E. S., and M. S. Bergman. 2000. Electrostatic N95 respirator filter media
efficiency degradation resulting from intermittent sodium chloride aerosol
exposure. Applied Occupational and Environmental Hygiene 15(8):600-608.
Myers, W.R., H. Kim, and N. Kadrichu. 1991. Effect of particle size on assess-
ment of faceseal leakage. Journal of the International Society for Respiratory
Protection 9:6-21.
Nelson, T. J., and C. E. Colton. 2000. The effect of inhalation resistance on
facepiece leakage. Aihaj 61(1):102-105.
Nicas, M. 1995. Respiratory protection and the risk of Mycobacterium tubercu-
losis infection. American Journal of Industrial Medicine 27(3):317-333.
NIOSH (National Institute for Occupational Safety and Health). 2004a. 42 Code
of Federal Regulations, Part 84. Respiratory Protective Devices. Federal Reg-
ister 42(1):505-596.
NIOSH. 2004b. Program concept for total inward leakage (TIL) performance
requirements and test methods. http://www.cdc.gov/niosh/npptl/standardsdev/
til/ (accessed June 20, 2007).
OCR for page 77
111
DESIGNING AND ENGINEERING EFFECTIVE PPE
NIOSH. 2006. Proposed industrial powered, air-purifying respirator (PAPR)
standard. http://www.cdc.gov/niosh/npptl/standardsdev/other/papr/concepts
/pdfs/paprcon091906.pdf (accessed June 20, 2007).
OSHA (Occupational Safety and Health Administration). 1998. 1910.134 Respi-
ratory protection. Federal Register 63(5):1270-1300.
OSHA. 2001. Bloodborne pathogens. Occupational Safety and Health Standards
(29 CFR Part 1910.1030). http://www.osha.gov/pls/oshaweb/owadisp.show_
document?p_table=STANDARDS&p_id=10051 (accessed June 26, 2007).
OSHA. 2006. Assigned protection factors. Federal Register 71(164):50121-
50192.
OSHA. 2007a. Air-purifying vs. atmosphere-supplying respirators. http://
www.osha.gov/SLTC/etools/respiratory/air_pure_vs_atmos_supp/air_pure_vs
_atmos_supp.html (accessed June 20, 2007).
OSHA. 2007b. Pandemic influenza preparedness and response guidance for
healthcare workers and healthcare employers. OSHA 3328-05. Washington,
DC.
Park, S., and S. Jayaraman. 2003. Enhancing the quality of life through wearable
technology. IEEE Engineering in Medicine and Biology Magazine 22(3):
41-48.
Pasko, S. 2007. Designing and engineering PPE: Next steps. Presentation at the
Institute of Medicine Workshop on Personal Protective Equipment for
Healthcare Workers in the Event of Pandemic Influenza: Next Steps and Re-
search Directions, February 22, 2007. Washington, DC.
Pissiotis, C. A., V. Komborozos, C. Papoutsi, and G. Skrekas. 1997. Factors that
influence the effectiveness of surgical gowns in the operating theatre. Euro-
pean Journal of Surgery 163(8):597-604.
Qian, Y., K. Willeke, S. A. Grinshpun, J. Donnelly, and C. C. Coffey. 1998.
Performance of N95 respirators: Filtration efficiency for airborne microbial
and inert particles. American Industrial Hygiene Association Journal
59(2):128-132.
Radonovich, L. 2007. Tolerability of commonly used respirators in healthcare
workers. Presentation at the Institute of Medicine Workshop on Personal Pro-
tective Equipment for Healthcare Workers in the Event of Pandemic Influ-
enza: Next Steps and Research Directions. February 22, 2007. Washington,
DC.
Rajamanickam, R., S. Park, and S. Jayaraman. 1998. A structured methodology
for the design and development of textile structures in a concurrent engineer-
ing environment. Journal of the Textile Institute 89(3):44-62.
Rosen, H. R. 1997. Acquisition of hepatitis C by a conjunctival splash. Ameri-
can Journal of Infection Control 25(3):242-247.
Rutala, W. A., and D. J. Weber. 2001. A review of single-use and reusable
gowns and drapes in health care. Infection Control and Hospital Epidemiol-
ogy 22(4):248-257.
OCR for page 77
112 PREPARING FOR AN INFLUENZA PANDEMIC
Ryan, M. A., R. S. Christian, and J. Wohlrabe. 2001. Handwashing and respira-
tory illness among young adults in military training. American Journal of
Preventive Medicine 21(2):79-83.
Siegel, J. D., E. Rhinehart, M. Jackson, L. Chiarello, and the Healthcare Infec-
tion Control Practices Advisory Committee. 2007. Guideline for isolation
precautions: Preventing transmission of infectious agents in healthcare set-
tings, 2007. http://www.cdc.gov/ncidod/dhqp/gl_isolation.html (accessed July
31, 2007).
Smith, J. W., and R. L. Nichols. 1991. Barrier efficiency of surgical gowns. Are
we really protected from our patients’ pathogens? Archives of Surgery
126(6):756-763.
Sun, G., and X. Xu. 1998. Durable and regenerable antibacterial finishing of
fabrics: Biocidal properties. Textile Chemist and Colorist 30(6):26-30.
Teleman, M. D., I. C. Boudville, B. H. Heng, D. Zhu, and Y. S. Leo. 2004. Fac-
tors associated with transmission of severe acute respiratory syndrome among
health-care workers in Singapore. Epidemiology and Infection 132(5):
797-803.
Treanor, J. 2005. Influenza virus. In Mandell, Douglas and Bennett’s principles
and practice of infectious diseases, 6th ed., edited by G. L. Mandell, J. E.
Bennett, and R. Dolin. New York: Elsevier Churchill Livingstone. Pp. 2060-
2085.
White, C., R. Kolble, R. Carlson, N. Lipson, M. Dolan, Y. Ali, and M. Cline.
2003. The effect of hand hygiene on illness rate among students in university
residence halls. American Journal of Infection Control 31(6):364-370.
Yassi, A., E. Bryce, D. Moore, R. Janssen, R. Copes, K. Bartlett, M. Fitzgerald,
M. Gilbert, P. Bigelow, Q. Danyluk, et al. 2004. Protecting the faces of
health care workers: Knowledge gaps and research priorities for effective
protection against occupationally-acquired respiratory infectious diseases.
British Columbia, Canada: The Change Foundation. http://www.cher.ubc.ca/
PDFs/Protecting_Faces_Final_Report.pdf (accessed June 11, 2007).