... Appendixes
From page 35...
... The informing power, Pint, of a measuring device is the number of bits required to encode the information potentially available from the device. Supposing, for example, that the device can report one of S possible values, the informing power of the device is log2~5~.
From page 36...
... Asbury and Hill report that for a typical IMS crevice, the number of theoretical plates rarely exceeds N = 5,000, which corresponds to R = 30. The number of intensity values that can be reported, S
From page 37...
... in aviation security reports a range of approximately 0 to 15 ms. The literature indicates considerable variation in the drift times observed, clepencling (as one would expect)